	* Basic parameter: 
  1)- Dimensions (mm): 140(L)×150(W)×220(H) 
  2)- Weight: about 1.8kg 
  3)- Ambient temperature: +5℃~43℃  
  4)- Relative humidity: ≤90%RH (Under without condensation status) 
  5)- Power supply: AC 90V~AC 242V, 50HZ 
  6)- Built-in battery: DC 12V 1800mAh rechargeable Ni-MH battery, the battery can be used continuously for 2 hours after full charged (24hours) under 25ml/h flow rate. 
  7)- Atmospheric pressure: 550hPa~1060hPa 
  8)- Storage temperature: -40℃~+55℃ 
  9)- Storage humidity: ≤95%RH 
  10)- Power consumption: ≤10VA 
  11)- Flow rate options: 1ml/h~1200ml/h, in 1ml/h increment 
  12)- Infusion total amount setting: 1ml~9999ml, in 1ml increment 
  13)- Occlusion pressure level: High:  825mmHg ± 150 mmHg (110kPa±20kPa) 
                            Middle:600mmHg ± 225mmHg (80kPa±30kPa) 
                            Low:  450mmHg ± 150 mmHg (60kPa±20kPa) 
  14)- Flow rate error: ±5% 
  15)- Alarm: occlusion, finish, low battery, standby (no operation), air in line, door open, motor troubles, AC power disconnected, starting up by battery 
  16)- Function: self-test, infusion amount setting, flow rate and infused volume display, power indication, safe flow rate limit, data preserves, KVO (Keep Vein Open) function 
* MOQ: 12 pcs 
* HS code: 9018909010 
* Packing: 1 pc/carton, carton size:360x215x235mm, carton G.w./N.w.:2.5/1.8 kgs. 
                                         | 
                                       
                                    |