* Basic Parameter:
1)- Dimensions(mm): 140(L)×160(W)×220(H)
2)- Weight: about 2.15Kg
3)- Power Supply:AC 90V~AC 242V、50HZ
4)- Built-in batteries: DC 12V 1800mAh
5)- Battery Model:JBT Ni-MH
6)- Battery Continuous Use Time:Above 6hrs
7)- Power consumption:≤18VA
8)- Infusion Flow Selection: 0.1ml/h~1200ml/h(6~150d/min), original value is 0.1ml/h (1d/min)
9)- Infusion Range:0.1ml~9999.9ml,original value is 0.1ml
10)- Flow Rate Precision:volume mode ±5%,drop mode ±2d
11)- Occlusion Pressure Level:High:110kPa±20kPa (825mmHg±150 mmHg)
Middle:80kPa±30kPa (600mmHg±225mmHg)
Low:60kPa±20kPa (450mmHg±150 mmHg)
12)- Use Environment:Temperature 5℃~43℃
13)- Relative Humidity ≤90%RH (Under without condensation status)
14)- Atmospheric Pressure 860hPa~1060hPa
15)- Storage Environment:Temperature -40℃~+55℃
16)- Relative Humidity ≤95%RH (Under without condensation status)
17)- Atmospheric Pressure 860hPa~1060hPa
18)- Waterproof Level: IPX1 (drip-proof)
19)- Electric Shock Protection: Class I, Type BF
20)- Alarm: occlusion, finish, low battery, standby (no operation), air in line, door open, AC power disconnected.
* Special Functions:
21)- Keep-Vein-Open (KVO): After finishing predetermined infusion volume, to prevent the formation of thrombus, will continue infusion with agreed minimum rate (3ml/h or 1ml/h).
22)- Power Supply Switching Function: When the AC power supply is off, it can be automatically switched to built-in battery power supply and prompt at the same time. When AC is switched on, automatically converted back to AC supply.
23)- Repeat Alarm Function: After the operator presses the “SILENCE” key, if the reason causing the alarm has been untreated for 2 minutes, the alarm will sound again to alert the operator to solve the problem.
24)- Data Memory Function: Save the last setting of sound volume level, flow precision fine tuning level and occlusion pressure level, as well as the clock, to prepare for the next use. The longest memory time is 5 years.
25)- High Flow Safety Confirm: In order to ensure infusion safety, when the flow rate is greater than or equal to 300ml/h (or equivalent), the device will prompt user to confirm again, and will not carry out infusion before confirmation.
26)- Light: For convenience, when installing or removing IV set at night or in darkness, light will automatically turn on after the door is open. The light is an LED light source (white).
27)- Battery Protection Alarm: In order to protect the battery and ensure infusion safety, there are different levels of acousto-optic prompts when the electric quantity of the battery is changing from close to depletion to almost depleted.
28)- Bubble Alarm: A continuously large number of bubbles going into the human body can form the air embolus, and it is life-threatening. So when bubble length is 3mm (10.6ul) or accumulated bubble length is 3mm (10.6ul) is detected, will trigger acoustic-optic alarm.
29)- Drip Sensor Slope: In d/min mode, when there are problems such as tube slope, liquid overflow and misplaced/moved sensor, there will be an alarm.
* MOQ: 12 pcs
* HS code: 9018909010
* Packing: 1 pc/carton, carton size:360x215x235mm, carton G.w./N.w.:2.8/2.15 kgs. |
|